Abstract

PurposeThe main objective of this study was to develop a technique to accurately determine the air gap between the end of the proton beam compensator and the body of the patient in proton radiotherapy.MethodsOrthogonal x‐ray image‐based automatic coordinate reconstruction was used to determine the air gap between the patient body surface contour and the end of beam nozzle in proton radiotherapy. To be able to clearly identify the patient body surface contour on the orthogonal images, a radiopaque wire was placed on the skin surface of the patient as a surrogate. In order to validate this method, a Rando® head phantom was scanned and five proton plans were generated on a Mevion S250 Proton machine with various air gaps in Varian Eclipse Treatment Planning Systems (TPS). When setting up the phantom in a treatment room, a solder wire was placed on the surface of the phantom closest to the beam nozzle with the knowledge of the beam geometry in the plan. After the phantom positioning was verified using orthogonal kV imaging, the last pair of setup kV images was used to segment the solder wire and the in‐room coordinates of the wire were reconstructed using a back‐projection algorithm. Using the wire as a surrogate of the body surface, we calculated the air gaps by finding the minimum distance between the reconstructed wire and the end of the compensator. The methodology was also verified and validated on clinical cases.ResultsOn the phantom study, the air gap values derived with the automatic reconstruction method were found to be within 1.1 mm difference from the planned values for proton beams with air gaps of 85.0, 100.0, 150.0, 180.0, and 200.0 mm. The reconstruction technique determined air gaps for a patient in two clinical treatment sessions were 38.4 and 41.8 mm, respectively, for a 40 mm planned air gap, and confirmed by manual measurements. There was strong agreement between the calculated values and the automatically measured values, and between the automatically and manually measured values.ConclusionsAn image‐based automatic method has been developed to conveniently determine the air gap of a proton beam, directly using the orthogonal images for patient positioning without adding additional imaging dose to the patient. The method provides an objective, accurate, and efficient way to confirm the target depth at treatment to ensure desired target coverage and normal tissue sparing.

Highlights

  • Delivering correct amount of dosage to the tumor target without affecting surrounding healthy tissue is an important mission in radiation oncology.[1]

  • Since the air gap is measured from the patient skin surface and it is normally difficult to clearly identify soft tissue on x‐ray images, a radiopaque wire can be placed along the patient skin surface during the patient treatment setup in the general location where the proton beam is aimed at

  • We have developed an orthogonal kV imaging‐based method to automatically reconstruct a wire from the digital images and to derive the proton beam air gaps at treatment

Read more

Summary

Introduction

Delivering correct amount of dosage to the tumor target without affecting surrounding healthy tissue is an important mission in radiation oncology.[1]. With the exception of limited number of proton centers equipped with three‐dimensional (3D) imaging positioning systems,[8] the two‐dimensional (2D) orthogonal imaging‐based system is still widely employed at many proton centers for the purpose of patient positioning. Unlike the 3D imaging system which can directly provide target depth information, with the 2D imaging system, the target depth information can only be indirectly derived by measuring the air gap between patient surface and beam nozzle prior to treatment. It would be desirable to utilize the same imaging process for patient treatment position purpose to determine the air gap, at no expense of additional imaging time and dose, even for the proton machines equipped with only two‐dimensional kilo‐voltage (kV) imaging systems. The goal of this study is to develop an orthogonal imaging‐ based automatic method for the determination of the air gap at proton therapy treatment

Objectives
Methods
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.