Abstract
A new projection reconstruction technique utilizes the oversampling of low spatial frequencies to estimate and correct for off-resonance effects. Interleaved spokes are acquired at one of two different echo times. From separated early-TE and late-TE raw data, two one-quarter resolution images are reconstructed and a one-quarter resolution field map is computed. Multifrequency reconstruction with all the data is then used to simultaneously correct for off-resonance and compensate for the difference in echo times. Resulting images obtained on phantoms and in vivo demonstrate significantly reduced off-resonance artifact without the acquisition of a separate field map.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.