Abstract

Automatic adjustment of only the spark-gap width in a spark-source mass spectrometer does not ensure that optimum conditions of electrode geometry are maintained with respect to the ion-optics system. A device has been developed which simultaneously maintains a constant gap width and also a more constant z-axis ion-illumination angle. This is the first development to utilize ion-optics parameters to adjust the sparking electrodes automatically. The system maintains the electrodes in an optimum configuration such that higher and more constant instrument sensitivity is maintained automatically. In addition, a significant improvement in the precision of instrumental response is demonstrated. It appears that relative isotopic abundances can be determined directly by the spark-source method which are comparable to those obtained in some cases by surface ionization or by electron bombardment. Results are given that support this contention.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.