Abstract

Guidelines for managing hyperbilirubinemia in newborns were developed by using diazo methods that measure total and direct-reacting bilirubins and calculate an indirect fraction. The automated Kodak Ektachem system allows for measuring serum bilirubin by either of two dry-slide methods: TBIL, involving a modified diazo method, and NBIL, involving a dual-wavelength colorimetric method that fractionates and directly measures the unconjugated (Bu) and conjugated (Bc) bilirubins (Bu+Bc = neonatal bilirubin). The manufacturer recommends that NBIL be used in newborns less than 15 days old, which is impractical in a large, busy hospital laboratory. We compared NBIL and TBIL in 500 paired serum samples from infants less than 15 days old. We noted a statistically significant difference between TBIL and NBIL values (162.9, SD 70.4, vs 164.6, SD 69.2, mumol/L; P less than 0.0001), which was small and of no clinical significance. We conclude that TBIL values may be used with caution for newborn bilirubin screening. Furthermore, NBIL measurements are an acceptable alternative to diazo measurements for neonatal care, allowing the use of previously developed guidelines with NBIL values.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.