Abstract

To determine the consistency of bilirubin values between the direct spectrophotometric (DS) and colorimetric diazo (diazo) methods in neonatal venous plasma specimens. We measured the total bilirubin via the DS and diazo methods in 255 neonatal venous plasma specimens and compared the overall and subgroup results. Slight underestimation of total bilirubin values in most specimens using the DS method was observed, with higher mean biases found in higher concentrations. Significantly high positive correlations were found in all groups in which most of the different values were within the limits of agreement. DS cutoff of > 12 mg/dL showed 100% for all predictive indices in comparison with the diazo cutoff > 15 mg/dL. Measurement of total bilirubin in neonatal venous plasma using the DS method had favorable agreement and high correlation with the diazo method. Therefore, the direct spectrophotometric method can be used as a reliable screening method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.