Abstract

Fluorescence Resonance Energy Transfer (FRET) microscopy has been widely used to study the structure and dynamics of molecules of biological interest, such as nucleic acids and proteins. Single molecule FRET (sm-FRET) measurements on immobilized molecules permit long observations of the system -effectively until both dyes photobleach- resulting in time-traces that report on biomolecular dynamics with a broad range of timescales from milliseconds to minutes. To facilitate the acquisition of large number of traces for statistical analyses, the process must be automated and the sample environment should be tightly controlled over the entire measurement time (~12 hours). This is accomplished using an automated scanning confocal microscope that allows the interrogation of thousands of single molecules overnight, and a microfluidic cell that permits the controlled exchange of buffer, with restricted oxygen content and maintains a constant temperature throughout the entire measuring period. Here we show how to assemble the microfluidic device and how to activate its surface for DNA immobilization. Then we explain how to prepare a buffer to maximize the photostability and lifetime of the fluorophores. Finally, we show the steps involved in preparing the setup for the automated acquisition of time-resolved single molecule FRET traces of DNA molecules.