Abstract
A bioanalytical method for the determination of atovaquone in plasma and whole blood by solid-phase extraction and high-performance liquid chromatography has been developed and validated. A structurally similar internal standard was added and protein was precipitated from plasma and whole blood with acetonitrile before being loaded on to a C 8 solid-phase extraction column. Atovaquone and internal standard were analysed by high-performance liquid chromatography on a C 18 J’Sphere ODS-M80 (150×4.0 mm) column with mobile phase acetonitrile–phosphate buffer, 0.01 M, pH 7.0 (65:35, v/v) and UV detection at 277 nm. The intra-assay precisions for plasma and whole blood were 2.2% and 1.9% respectively at 12 μ M and 6.0% and 5.6% respectively at 0.75 μ M. The inter-assay precisions for plasma and whole blood were 1.4% and 2.1% respectively at 12 μ M and 4.9% and 3.4% respectively at 0.75 μ M. The lower limit of quantification in plasma and whole blood were 150 n M. The limit of detection in plasma and whole blood were 30 n M.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.