Abstract
In the first part of this work (see preceding paper) the theory of the method lying upon a novel concept of birefringence, azimuths, and transmission mapping in large area (up to 6 in. diameter) wafers has been presented. The arrangement consisting of two HR-type linear polarizers rotated simultaneously by a stepper motor versus an immobile wafer and using a video frame grabber (VFG)/TV camera detecting system is capable of collecting data and plotting the three maps within a fraction of a minute. A detailed error analysis presented in the preceding paper has shown that in usual circumstances the VFG with 256 grey levels enables determination of birefringence with an error not greater than approximately 5×10−7, whereas errors of the azimuths and transmission are fractions of a degree and of a percent, respectively. In this part of the work the arrangement constructed is fully described and a set of polariscopic images and measured maps are presented for an exemplary 4 in. GaAs wafer.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.