Abstract
PACIA, a homogeneous non-radioimmunoassay, has been adapted to the determination of serum alpha 1-fetoprotein. This technique is based on the agglutination of latex particles coated with antibodies to the antigen to be determined. The agglutination is measured by using an optical cell counter designed to count blood cells, to determine the reduction in the number of non-agglutinated particles. Interferences by serum constituents are avoided by coating the particles with the F(ab')2-fragments of th immunoglobulin G fraction of the antiserum. The system is automated, with a sampling rate of 50/h and an incubation time of 26 min. Concentrations used in preparing the standard curve ranged from 1 to 50 microgram/L; analytical recoveries were 93.5 to 98.4%; the correlation coefficient of PACIA with radioimmunoassay, calculated from results on 127 samples, was 0.98; maximum within- and between-assay CVs were 7.4% and 9.6%, respectively.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.