Abstract

We describe a non-radioisotopic automated immunoassay for serum digoxin in the therapeutic range at 50 samples per hour by use of the particle-counting immunoassay technique and a special reagent, rheumatoid factor. Digoxin is determined by evaluating its inhibitory activity on the agglutination of digoxin-coated latex particles. The agglutinating system consists of the anti-digoxin IgG mixed with rheumatoid factor. This anti-IgG autoantibody reacts with IgG when they are aggregated or bound to a surface. Interference by serum proteins was prevented by pepsin digestion at 37 degrees C. Correlation with radioimmunoassay was r = 0.94 for 109 samples with concentrations between 0 and 5.7 micrograms/L. Analytical recovery overall averaged 102%. Within-assay precision (CV) ranged from 12.5% at 0.96 micrograms/L to 4.4% at 2.97 micrograms/L; between-assay precision varied from 8.4% at 0.75 micrograms/L to 5.8% at 2.92 micrograms/L.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.