Abstract
We describe automation of liquid injection field desorption/ionization (LIFDI) for reproducible sample application, improved spectral quality, and high-throughput analyses. A commercial autosampler provides reproducible and unattended sample application. A custom-built field desorption (FD) controller allows data station or front panel control of source parameters including high-voltage limit/ramp rate, emitter heating current limit/ramp rate, and feedback control of emitter heating current based on ion current measurement. Automated LIFDI facilitates ensemble averaging of hundreds of Fourier transform ion cyclotron resonance mass spectra for increased dynamic range, mass accuracy, and S/N ratio relative to single-application FD experiments, as shown here for a South American crude oil. This configuration can be adapted to any mass spectrometer with an LIFDI probe.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.