Abstract

Markerless patient registration based on the facial skin surface makes logistics prior to image-guided surgery much easier, as it is not necessary to place and measure registration markers. A laser scan registration of the surgical site takes the place of conventional marker-based registration. In a clinical study, the stability and accuracy of markerless patient registration was evaluated in 12 patients. Intraoral titanium markers served as targets for the infrared-pointer of the navigation system in order to check the accuracy of the markerless registration process. The correlation between laser scan resolution and navigation accuracy was checked using seven different laser scan resolutions (a cloud of 300,000 laser scan points down to 3750 laser scan points of the surgical site). The markerless patient registration was successful as long as high laser scan resolution was used (30,000 laser scan points and more): the titanium markers were detected with a mean deviation of 1.1±0.2 mm. Low resolution laser scans (6000 laser scan points of the surgical site and less) revealed inaccuracies up to 6 mm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.