Abstract

The purpose of this study was to determine if the Advia 2120 hematology analyzer (Siemens Healthcare Diagnostics, Deerfield, IL) can accurately quantify both WBCs and the proportion of neutrophils from synovial and ascitic body fluids. We analyzed 60 samples on the Advia and compared the results with manual counts and smear reviews. We also assessed the effect of adding hyaluronidase to the samples. WBC counts and the proportion of neutrophils reported by the hematology analyzer were harmonized and highly correlated with manual counts and fluid smear reviews. The addition of hyaluronidase to the synovial fluid consistently increased the WBC counts on both manual and automated analysis (P < .001). We conclude that the Advia hematology analyzer can be used for WBC and neutrophil counting of cells in synovial and ascitic fluids. Hyaluronidase should be added before manual or automated counting of cells in synovial fluids.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.