Abstract

The first commercially available centrifugal analyzer having fluorescence detection capability was used to develop kinetic fluorometric assays for several proteinases. The substrates were all synthetic oligopeptides incorporating 7-amino-4-trifluoromethylcoumarin, a molecule that can be detected either spectrophotometrically or fluorometrically. We thus compared the centrifugal analyzer spectrophotometric and fluorometric detection systems, finding fluorescence detection to be 50-fold more sensitive. We also compared the sensitivity of the fluorescence detector to that of a conventional spectrofluorometer by determining the minimum detection limit for each enzyme on both instruments; we found them to be similar in sensitivity. As an illustrative application, we measured the cathepsin B-like activity in serum samples from 55 women. The median enzyme activity of women taking oral contraceptives and pregnant women was increased two- and threefold, respectively, over the control group (about 5% CV within run, and 10% CV between runs).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.