Abstract

An extractionless method for determining aflatoxin M1 (AFM1), a major metabolite of aflatoxin B1 (AFB1), in human urine was developed. The biological fluid is injected directly into the chromatographic system after simple dilution and centrifugation. A pre-column, packed with a cation-exchange phase and coupled on-line to a column-switching liquid chromatography (LC) system, is used for sample pre-treatment and concentration. The analytes are non-selectively desorbed with the LC eluent and cleaned by means of a column-switching procedure. Pre-treatment and analysis were performed within 40 min. Average AFM1 recovery reached 97% in the 10–100 ng/l range of urine. The detection limit of AFM1 in urine and milk was 2.5 ng/l for 1 ml of injected sample. A comparison with an immunoaffinity column clean-up and LC method was performed. The method was applied to determine AFM1 in the urine of AFB1 gavaged rats, and in the urine of both potentially exposed and supposedly unexposed workers. The method was also extended to milk.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call