Abstract

A method for multiresidue pesticide analysis in food is described. After a conventional clean-up, gas chromatographic analysis is performed in a gas chromatograph equipped with two fused-silica capillary columns coated with methylsilicone SP 2100 and methylphenylsilicone OV-17. The effluent from each column is split to electron-capture and nitrogen—phosphorus detectors, which are connected to a dual channel integrator. Therefore, from each gas chromatographic run parallel records of signals from the two detectors are obtained. Calibration of the system is carried out for the SP 2100 column with three test mixtures covering all pesticides. Additionally, four internal standards are included, two responding to the electron-capture detector and the other two to the nitrogen—phosphorus detector. Automated analysis is performed with test mixtures and food samples on the SP 2100 column overnight as a screening procedure. After selection of positive samples a confirmatory test and quantitation are carried out manually applying appropriate test mixtures according to the results of the screening runs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.