Abstract

An automated HPLC method with column switching is described for the determination of quetiapine, clozapine, perazine, olanzapine and metabolites in blood serum. After clean-up on silica C8 material (20 microm particle size) drugs were separated on ODS Hypersil C18 material (5 microm; column size 250 mm x 4.6 mm i.d.) within 25 min and quantified by ultraviolet (UV) detection at 254 nm. The limit of quantification ranged between 10 and 50 ng/ml. At therapeutic concentrations of the drugs, the inter-assay reproducibility was below 10%. Analyses of drug concentrations in serum of 75-295 patients treated with therapeutic doses of the antipsychotic drugs revealed mean+/-S.D. steady state concentrations of 139+/-136 ng/ml for quetiapine, 328+/-195 ng/ml for clozapine, 48+/-27 ng/ml for olanzapine and 71+/-52 ng/ml for perazine. The method was thus suitable for routine therapeutic drug monitoring and may be extended to other drugs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.