Abstract
Bioanalytical methods based on automated solid-phase extraction (SPE) and high-performance liquid chromatography with electrospray tandem mass spectrometry (LC-MS-MS) have been developed and utilized for the determination of MMP inhibitors in plasma and cartilage tissues. The SPE methods were automated using a 96-well extraction plate and a 96-channel programmable liquid-handling workstation. The LC-MS-MS methods were developed using a rapid gradient LC separation, followed by sample introduction through an ionspray interface in the positive ion mode and tandem mass spectrometric detection with selected reaction monitoring. In the optimized SPE methods, crude plasma or ground cartilage supernatant samples were loaded onto an SPE plate to remove proteins and other interfering components in the matrixes to render relatively clean extracts for LC-MS-MS analysis. Compared to the simple plasma protein precipitation method, the automated SPE method afforded significant time-saving in sample preparation and improved sensitivity in MS detection. The methods were validated and successfully applied to the analysis of protease inhibitors in plasma and cartilage tissues.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.