Abstract

Flow disturbances in tapered arterial grafts of angles of taper between 0.5 and 1.0 deg were measured in vitro using a pulsed ultrasound Doppler velocimeter. The increase in transition Reynolds numbers with angle of taper and axial distance was determined for steady flow. The instantaneous centerline velocities were measured distal to a 50 percent area stenosis (as a model of a proximal anastomosis), in steady and pulsatile flow, from which the disturbance intensities were calculated. A significant reduction in post-stenotic disturbance intensity was recorded in the tapered grafts, relative to a conventional cylindrical graft. In pulsatile flow with a large backflow component, however, there was an increase in disturbance intensity due to diverging flow during flow reversal. This was observed only in the 1.0 deg tapered graft. These findings indicate that taper is an important consideration in the design of vascular prostheses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.