Abstract

An atrial pacing threshold is shown ( Figure 1A ). Modality of pacing is AAI at 90 bpm. The first paced beat is followed by two consecutive ventricular depolarizations. The second atrial paced beat is not followed by any ventricular-related sensed beat. The third is normally conduced to the ventricles. The electrocardiographic behaviour observed concomitantly to the first two atrial paced beats is explained with a double response ( Figure 1B ) (antegrade conduction over both the fast and the slow AV-nodal pathway) followed by a pseudo-atrioventricular block due to collision of the following antegrade impulse generated by atrial …

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.