Abstract

Atomic surface structure imaging is instrumental for the understanding of surface-related phenomena. Here, we show that conventional tapping mode atomic force microscopy with high cantilever eigenmodes and subnanometer amplitudes allow routine atomic imaging at atmospheric pressures. We identify the reasons for failure of atomic resolution imaging employing low eigenmodes. Strong tip-surface interactions cause significant differences between the oscillatory behaviors of the inclination of the cantilever as detected by conventional instruments and of the vertical position of the tip, which prevents correct functioning of instrumental feedback control loops. However, high effective spring constants of high eigenmodes make it possible to overcome the problem. Furthermore, the combination of high effective elastic constants of high cantilever eigenmodes with the high flexibility of the cantilever substantially enhances the imaging stability, thereby universally allowing atomic imaging of solid surfaces in gaseous environments and at elevated temperatures. Demonstrated imaging examples include single sulfur vacancies at the surface of ${\mathrm{MoS}}_{2}$ crystals imaged at temperatures ranging from room temperature to 250\ifmmode^\circ\else\textdegree\fi{}C and potassium ions on hydrophilic and highly adhesive muscovite mica surfaces. Moreover, the high imaging stability allows knocking atoms off the ${\mathrm{MoS}}_{2}$ surface by hard tapping, indicating the potential for ultrahigh resolution lithography.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call