Abstract

For high-resolution imaging without bulky external vibration isolation, this paper presents an atomic force microscope (AFM) capable of vibration isolation with its internal Z-axis (vertical) actuators moving the AFM probe. Lorentz actuators (voice coil actuators) are used for the Z-axis actuation, and flexures guiding the motion are designed to have a low stiffness between the mover and the base. The low stiffness enables a large Z-axis actuation of more than 700 µm and mechanically isolates the probe from floor vibrations at high frequencies. To reject the residual vibrations, the probe tracks the sample by using a displacement sensor for feedback control. Unlike conventional AFMs, the Z-axis actuation attains a closed-loop control bandwidth that is 35 times higher than the first mechanical resonant frequency. The closed-loop AFM system has robustness against the flexures’ nonlinearity and uses the first resonance for better sample tracking. For further improvement, feedforward control with a vibration sensor is combined, and the resulting system rejects 98.4% of vibrations by turning on the controllers. The AFM system is demonstrated by successful AFM imaging in a vibrational environment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.