Abstract

A calibration method is presented for determining the spring constant of atomic force microscope (AFM) cantilevers, which is a modification of the established Cleveland added mass technique. A focused ion beam (FIB) is used to remove a well-defined volume from a cantilever with known density, substantially reducing the uncertainty usually present in the added mass method. The technique can be applied to any type of AFM cantilever; but for the lowest uncertainty it is best applied to silicon cantilevers with spring constants above 0.7 N m−1, where uncertainty is demonstrated to be typically between 7 and 10%. Despite the removal of mass from the cantilever, the calibration method presented does not impair the probes’ ability to acquire data. The technique has been extensively tested in order to verify the underlying assumptions in the method. This method was compared to a number of other calibration methods and practical improvements to some of these techniques were developed, as well as important insights into the behavior of FIB modified cantilevers. These results will prove useful to research groups concerned with the application of microcantilevers to nanoscience, in particular for cases where maintaining pristine AFM tip condition is critical.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.