Abstract

Photon and neutron soft tissue absorbed doses near the entrance door to the medical LINAC treatment room were measured with the use of thermoluminescent detectors LiF:Mg,Cu,P in the anthropomorphic phantom. Two different therapeutic beams (6 and 15 MV) and four treatment techniques were involved in the present study. This allowed one to investigate the contribution of scattered X rays, secondary neutrons and gamma rays to the radiation field. Photon absorbed dose rates 50 cm away from the LINAC room door during emission of 15-MV X rays varied between 4.1×10(-4) and 5.6×10(-4) Gy h(-1), depending on the gantry position and the irradiation field size, whereas in the case of 6-MV therapeutic irradiation these doses are ∼1.5 times lower. In the case of 15-MV beam emission, a mixed radiation field near the bunker door is observed with the photon radiation as the main component, which includes a 33.1 % contribution of the induced gamma radioactivity and ∼2.1 % contribution of the neutron radiation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.