Abstract

This study aims to provide radiation reference levels in orthopaedic surgery. A total of 753 procedures were collected within 1y. Categories containing several similar procedures were created based on four criteria: same anatomical area, same level of complexity, only single procedures, and at least 10 cases per category. Exposure was defined in terms of air kerma-area product, fluoroscopy time, and air kerma at the patient entrance reference point. For common procedures, median effective doses to patient were calculated using the Monte Carlo Software PCXMC. Most irradiating procedure in this study i.e. intramedullary nailing of the proximal femur was equivalent to an air kerma at the patient entrance reference point of 37.1mGy, which is ~50 times lower than the threshold for acute deterministic effects of radiation. Optimization remains a must to reduce the dose while maintaining the image quality and reducing the likelihood of stochastic effects.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.