Abstract

The aim of this study is to evaluate the personal dose equivalent Hp(10) in the most frequent (non-cardiac) paediatric interventional radiology (PIR) procedures: central venous catheters (CVC), hepatic/biliary and sclerotherapy interventions. i2 active solid-state dosemeters placed over the lead apron were used to monitor the exposure of three interventional radiologists over 18months. A database was created to register all procedures performed by each radiologist (including the type of procedure and the kerma-area product, PKA). The mean Hp(10) per procedure for CVC, sclerotherapy and hepatic/biliary interventions was respectively 0.01±0.01mSv, 0.18±0.13mSv and 0.12±0.06mSv (k=2). A similar value of Hp(10)/PKA was found despite the type of procedure or the patient weight (~10μSv/Gy·cm2). There was high variability among individual interventions, probably due to the variable level of complexity, which led to uncertainties in the measurements' mean higher than those associated with the dosemeter's angular and energy dependence. i2 therefore proved suitable for monitoring Hp(10) in PIR procedures.