Abstract

A colorimetric loop-mediated isothermal amplification assay detects changes in pH during amplification based on color changes at a constant temperature. Currently, various studies havefocused on developing and assessing molecular point-of-care testing instruments. In this study, we evaluated amplified DNA concentrations measured using the colorimetric LAMP assay of the 1POT™ Professional device(1drop Inc, Korea). Results of the 1POTanalysis of clinical samples were compared with measurements obtained from the Qubit™ 4 and NanoDrop™ 2000 devices (both from Thermo Fisher Scientific, MA, USA). These results showed a correlation of 0.98 (95%CI: 0.96-0.99) and 0.96 (95% CI: 0.92-0.98) between 1POTand the Qubitand NanoDrop. 1POTcan measure amplified DNA accurately and is suitable for on-site molecular diagnostics.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.