Abstract

The aim of this study was to analyze the recommended OSEM image reconstruction parameters in positron emission tomography (PET). Spatial resolution, signal-to-noise ratio, and contrast were used as physical figures of merit (FOM). For statistical FOMs, the t-value and the area under the receiver operating characteristic (ROC) were employed. The spatial resolution was measured with 21 point sources. The signal-to-noise ratio, the contrast, and the t-value were investigated with a whole-body phantom with hollow spheres inserted. A phantom containing line sources was used for ROC analysis. As result, the reconstruction parameters recommended for visual evaluation lead to images with an adequate lesion detectability. The quantitative reconstruction, however, needs improvement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.