Abstract

In Brazil there are around 600 mammography equipment with CR system. Taking into account the quality of image is essential to evaluate the mean glandular dose so you can optimize the radiation protection of patients evaluated with this type of system. Therefore, this study aimed to determine the mean glandular dose and quality of image for the CR system of laboratory of radioprotection applied to mammography of the Centre of Development of Nuclear Technology (CDTN). For this, we evaluated the linearity of the detector’s response, contrast to noise ratio and signal to noise ratio, which, according to European protocol, showed results within acceptable limits. Next, evaluated the quality of image with the CDMAM Phantom and mean glandular dose to the detector Fluke Biomedical TNT 12000WD, where they presented, respectively, within the results expected by the manufacturer and the limits of acceptable and desirable by the European protocol. then, the CR system of CDTN is optimized.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.