Abstract

The objective of this study was to assess correlation between dimensions of ball phantom and distortion rate of panoramic radiography in dental cone beam computed tomography (CBCT) system using appropriate standard methods. The results revealed that only two measurement (2.99%) out of 67 measurements exceeded the recommended threshold value for distortion rate. There was no correlation between the ball diameter and the distortion produced. However, a low positive correlation was observed between the distortion rate and the ratio of the horizontal to the vertical diameter of the ball images. Also, the matlab code, presented results which are closer to the acceptable limit than the Romexis software. Patient positioning should be carefully and correctly considered to prevent and to bring the image distortion rate to its barest minimum for proper diagnosis in dental CBCT. For proper diagnosis and treatment planning, this distortion must be accounted for during clinical applications.