Abstract

Objective:The application of bubble-based ablation with the focus ultrasound therapy histotripsy is gaining traction for the treatment of venous thrombosis, among other pathologies. For extensive clot burden, the histotripsy source must be translated to ensure uniform bubble activity throughout the vascular obstruction. The purpose of this study was to evaluate the targeting accuracy of a histotripsy system comprised of a focused source, ultrasound image guidance, and a collaborative robot (cobot) positioner. The system was designed with a primary emphasis for treating deep vein thrombosis.Methods:Studies to test treatment planning and targeting bubble activity with the histotripsy-cobot system were conducted in an in vitro clot model. A tissue-mimicking phantom was also targeted with the system, and the predicted and actual areas of liquefaction were compared to gauge the spatial accuracy of ablation.Results:The system provided submillimeter accuracy for both tracking along an intended path (within 0.6 mm of a model vessel) and targeting bubble activity within the venous clot model (0.7 mm from the center of the clot). Good correlation was observed between the planned and actual liquefaction locations in the tissue phantom, with an average Dice similarity coefficient of 77.8%, and average Hausdorff distance of 1.6 mm.Conclusion:Cobots provide an effective means to apply histotripsy pulses over a treatment volume, with the ablation precision contingent on the quality of image guidance.Significance:Overall, these results demonstrate cobots can be used to guide histotripsy ablation for targets that extend beyond the natural focus of the transducer.

Highlights

  • H ISTOTRIPSY is an ablative form of focused ultrasound under development for the treatment of thromboocclusive disease [1], [2], among other clinical targets [3]

  • We examine a cobot-histotripsy system for the treatment of venous thrombosis ablation in vitro [1], [17], [18]

  • Studies to gauge the precision of the system in tracking a model vessel provide information regarding repeatability of the image guided, user executed system, and accuracy of the interpolation algorithm to translate the histotripsy focus along the intended treatment path

Read more

Summary

Introduction

H ISTOTRIPSY is an ablative form of focused ultrasound under development for the treatment of thromboocclusive disease [1], [2], among other clinical targets [3]. Hendley and Viktor Bollen are with the University of Chicago

Objectives
Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call