Abstract

Lawrence Berkeley National Laboratory developed a simple, efficient, and cost-effective replacement for the traditional glass column system used to monitor H and C emissions from rooftop stacks. The primary goals in developing a replacement (the modified jar system) were to 1) maintain or improve collection efficiency while keeping leakage to less than 5%, 2) simplify the set-up process, and 3) reduce costs. Both the traditional glass column assembly and the modified jar system were operated in tandem for a 13-mo period. Results showed that the modified sample jar system provided equivalent or improved collection efficiency for both H and C. Additional advantages included reduced leak-test errors, quicker and simpler set-up, and material costs that were reduced by nearly an order of magnitude.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.