Abstract

PurposeTo retrospectively assess the incidence and magnitude of air pockets around vaginal cylinders and its impact on dose distribution in vaginal cuff image-guided high-dose-rate (HDR) brachytherapy.Material and methodsFifty endometrial carcinoma patients treated by postoperative HDR vaginal cuff brachytherapy were included in the study. The average age of patients was 58.3 ± 11.8 years (range: 31-87 years). Brachytherapy was performed using cylindrical applicators, and the dose prescribed to 0.5 cm from the applicator's surface, over a length of 5 cm from the applicator's tip. Computed tomography (CT) simulation was used for each brachytherapy fraction. The incidence, vaginal mucosa displacement, volume, and dosimetric effect of air pockets around the vaginal cylinder were evaluated.ResultsA total of 78 air pockets were found in 29/50 patients (58%) and 45/135 (33%) brachytherapy plans. They were located at the apex: 16/78 (20%) and lateral to the applicator: 62/78 (80%). The volume of air pockets ranged between 0.01 and 2.1 cm3 (mean: 0.15 cm3 ± 0.36 cm3), and the maximum displacement of vaginal mucosa from cylinder surface was between 0.1 and 1.09 cm (mean: 0.34 cm ± 0.2 cm). The dose reduction to the vaginal mucosa generated by the air pockets ranged from 0.5 to 66% (mean: 26.4% ± 13.9%).ConclusionsThe presence of air pockets around vaginal cylinder applicators is frequently noticed in post-operative vaginal cuff brachytherapy. The dose to the vaginal mucosa is reduced, as a result of displacement generated by air pockets. The effect on the clinical outcome of this dose reduction is yet to be determined.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call