Abstract

This article documents the results of the first in a series of experiments designed to evaluate the suitability of a novel, high resolution, color, digital, liquid crystal display (LCD) panel for diagnostic quality, gray scale image display. The goal of this experiment was to measure the performance of the display, especially with respect to luminance. The panel evaluated was the IBM T221 22.2" backlit active matrix liquid crystal display (AMLCD) with native resolution of 3840 x 2400 pixels. Taking advantage of the color capabilities of the workstation, we were able to create a 256-entry grayscale calibration look-up table derived from a palette of 1786 nearly gray luminance values. We also constructed a 256-entry grayscale calibration look-up table derived from a palette of 256 true gray values for which the red, green, and blue values were equal. These calibrations will now be used in our evaluation of human contrast-detail perception on this LCD panel.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.