Abstract

Analytical scale active flow technology first generation silica monolithic columns kitted out in curtain flow mode of operation were studied for the first time. A series of tests were undertaken assessing the column efficiency, peak asymmetry and detection sensitivity. Two curtain flow columns were tested, one with a fixed outlet ratio of 10% through the central exit port, the other with 30%. Tests were carried out using a wide range in inlet flow segmentation ratios. The performance of the curtain flow columns were compared to a conventional monolithic column. The gain in theoretical plates achieved in the curtain flow mode of operation was as much as 130%, with almost Gaussian bands being obtained. Detection sensitivity increased by as much as 250% under optimal detection conditions. The permeability advantage of the monolithic structure together with the active flow technology makes it a priceless tool for high throughput, sensitive, low detection volume analyses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.