Container closure integrity (CCI) testing is required by different regulatory authorities in order to provide assurance of tightness of the container closure system against contamination, for example, by microorganisms. Microbial ingress CCI testing is performed by incubation of the container closure system with microorganisms under specified testing conditions. Physical CCI uses surrogate endpoints, such as coloration by dye solution ingress or gas flow. In order to correlate microbial ingress CCI and physical CCI test methods and to evaluate the methods' capability to detect a given leak, artificially created defects (artificial leaks) are being introduced into the container closure system in a variety of different ways. In our study, artificial leaks were generated using inserted copper wires between the glass vial opening and rubber stopper. Up to date, the insertion of copper wires introduced leaks of unknown size and shape. With nonlinear finite element simulations, the effective aperture size between the rubber stopper and the glass vial was calculated, depending on wire diameter and capping force, and the leak shape was modelled. Our results suggest that the size as well as the shape of the artificial leaks should be taken into account when evaluating critical leak sizes, as flow rate does not, independently, correlate to the hole size.