Abstract

Arthrodesis of the first metatarsophalangeal joint of 21 matched pairs of cadaver toes was performed in order to compare the strength of three methods of internal fixation: 1. two crossed cannulated screws, 2. a dorsal plate with an oblique 0.062 K-wire, and 3. two compression staples with an oblique 0.062 K-wire. Biomechanical testing with plantar force was carried out, and gapping across the fusion site was measured. Stiffness, load to 1-mm displacement, and force to failure was determined for each specimen. Both the plate and screw constructs were statistically stronger in force to failure and initial stiffness than the compression construct. Compression staples have an advantage in their ease of insertion and theoretical continuous compressive force across an arthrodesis site, but should be supplemented with a cast or other external immobilization until union is achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.