Abstract

Dynamic contrast-enhanced (DCE) MRI is often used to measure the transfer constant (Ktrans) and distribution volume (ve) in pelvic tumors. For optimal accuracy and reproducibility, one must quantify the arterial input function (AIF). Unfortunately, this is challenging due to inflow and signal saturation. A potential solution is to use MR signal phase (ϕ), which is relatively unaffected by these factors. We hypothesized that phase-derived AIFs (AIFϕ) would provide more reproducible Ktrans and ve values than magnitude-derived AIFs (AIF|S|). We tested this in 27 prostate dynamic contrast-enhanced MRI studies (echo time=2.56 ms, temporal resolution=13.5 s), using muscle as a standard. AIFϕ peak amplitude varied much less as a function of measurement location (inferior-superior) than AIF|S| (5.6±0.6 mM vs. 2.6±1.5 mM), likely as a result of ϕ inflow insensitivity. However, our main hypothesis was not confirmed. The best AIF|S| provided similar reproducibility versus AIFϕ (interpatient muscle Ktrans=0.039±0.021 min(-1) vs. 0.037±0.025 min(-1), ve=0.090±0.041 vs. 0.062±0.022, respectively).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.