Abstract

The measured response of rapid biochemical oxygen demand (BOD) biosensors is often not identical to those measured using the conventional 5-day BOD assay. This paper highlights the efficacy of using both glucose-glutamic acid (GGA) and Organisation for Economic Cooperation and Development (OECD) BOD standards as a rapid screen for microorganisms most likely to reliably predict real effluent BODs when used in rapid BOD devices. Using these two synthetic BOD standards, a microorganism was identified that produced comparable BOD response profiles for two assays, the MICREDOX® assay and the conventional 5-day BOD(5) test. A factorial experimental design systematically evaluated the impact of four factors (microbial strain, growth media composition, media strength, and microbial growth phase) on the BOD response profiles using GGA and OECD synthetic standard substrates. An outlier was identified that showed an improved correlation between the MICREDOX® BOD (BOD(sens)) and BOD(5) assays for both the synthetic standards and for real wastewater samples. Microbial strain was the dominant factor influencing BOD(sens) values, with Arthrobacter globiformis single cultures clearly demonstrating superior rapid BOD(sens) response profiles for both synthetic and real waste samples. It was the only microorganism to approach the BOD(5) response for the OECD substrate (171 mg O(2)L(-1)), and also reported BOD values for real waste samples that were comparable to those produced by the BOD(5) test, including discriminating between filtered and unfiltered samples.