Abstract

Wavelength-Modulated Differential Laser Photothermal Radiometry (WM-DPTR) has been designed for noninvasive glucose measurements in the mid-infrared (MIR) range. Glucose measurements in human blood serum in the physiological range (20-320 mg/dl) with predicted error <10.3 mg/dl demonstrated high sensitivity and accuracy to meet wide clinical detection requirements, ranging from hypoglycemia to hyperglycemia. The glucose sensitivity and specificity of WM-DPTR stem from the subtraction of the simultaneously measured signals from two excitation laser beams at wavelengths near the peak and the baseline of the strongest interference-free glucose absorption band in the MIR range. It was found that the serum glucose sensitivity and measurement precision strongly depend on the tunability and stability of the intensity ratio and the phase shift of the two laser beams. This level of accuracy was favorably compared to other MIR techniques. WM-DPTR has shown excellent potential to be developed into a clinically viable noninvasive glucose biosensor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.