Abstract

The guaranteeing approach is applied to the joint calibration problem of an accelerometer unit and a high-precision bench. An optimal calibration design with the minimum total number of angular positions of the bench is obtained. Optimal guaranteeing estimates of the requisite parameters are constructed. In practice, the necessary positions of the bench are often difficult to implement precisely; due to this fact, a procedure to take into account the inaccurate setup errors of the bench is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.