Abstract

Drift in signals limits the performance of inductively-coupled plasma (i.c.p.) emission spectrometry. The generalized internal reference method was applied to deconvolute the drift of an i.c.p. system. The plasma parameters were correlated to the drift components and the error boundary for the method was found. In the system studied, it was found that fluctuations in the nebulizer uptake rate were responsible for short-term signal changes. Long-term signal drift was caused mainly by flow variations of both cooling gas and aerosol carrier gas. The incident power to the plasma was found to be stable compared to other sources of drift. The method was found to be adequate for evaluating the origins of i.c.p. drift.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.