Abstract

2X magnification employing a 200-mu focal spot and an Alpha 8-XM screen/film system was applied to oral cholecystography and the results compared with those for the conventional contact technique with the Par-RP system. The basic imaging properties of the system, as well as phantom studies, indicated that the image quality obtained with magnification is comparable to or better than that for the conventional technique. In clinical studies on the detection of gallstones, the conventional technique revealed 5 true-positive and 17 true-negative cases and 1 false-positive and 2 false-negative cases, while the magnification technique provided 7 true-positive and 18 true-negative cases but no false cases. With the magnification technique the skin dose was reduced to approximately half that for the conventional contact technique.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.