Abstract
To evaluate three non-invasive assays for the diagnosis of schistosomiasis mansoni in an Egyptian village. Urine was collected for the detection of circulating cathodic antigen (CCA) and cell-free parasite DNA (cfpd) by Point-of-contact (POC)-cassette assay and PCR, respectively. These tests were compared to Kato-Katz (KK) faecal thick smear for detection of Schistosoma mansoni eggs. Disease prevalence by POC-CCA assay was 86%; by PCR it was 39% vs. 27% by KK. Compared to KK, the sensitivity of POC-CCA reached 100%, but its specificity was only 19.2% with 41% accuracy. Sensitivity of the PCR assay for cfpd was 55.56%, and specificity was 67.12% with 64% accuracy. A new end point was calculated for combined analysis of KK, POC-CCA assay and PCR. Sensitivity for the three tests was 52.94%, 90.2% and 76.47%; specificity was 100% for KK and PCR and 18.37% for POC-CCA. The accuracy calculated for the three tests at the end point was 76% for KK, 55% for POC-CCA assay and 88% for PCR. Conventional PCR assay for detection of cfpd provides a potential screening tool for intestinal schistosomiasis with reliable specificity, reasonable accuracy and affordable financial and technical cost.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.