Abstract
Abstract Proton-induced X-ray analysis was studied at the 5 MV Van de Graaff accelerator of Tohoku University with emphasis on backing materials and the angular dependence of the background due to bremsstrahlung. From the practical viewpoint, 4 μm Mylar films among Formvar, carbon and Mylar gave sufficiently low background spectra and good mechanical strength. The detection limit was much improved in the backward direction in comparison with that at 90° which has usually been adopted. A detection sensitivity on the subpicogram level was obtained for an extensive range of elements covering a wide range of atomic number using 4 μm Mylar film backing.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.