Abstract
Proton induced X-ray analysis has been studied by the 5 MV Van de Graaff of Tohoku University emphasizing on backing materials and angular dependence of background. The method has further been applied to analyses of some environmental samples. From experimental results, it was found that 4 μm Mylar foil among Formvar, Mylar and carbon gives sufficiently low background spectrum, high sensitivity and good mechanical strength. The detection limit is much improved at the backward direction with respect to the incident beam comparing with that at 90° which has usually been adopted. The analytical results on soil and aerosol samples were compared with those by the Atomic Absorption Spectrophotometrical method (AAS). Vanadium can easily be analysed by this method, but not by the AAS method. Milk and human milk were also analysed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.