Abstract
In this work we present a refractive index (RI) sensor based on a sub-wavelength holographic diffraction grating. The sensor chip was fabricated by dry etching of the finely spaced (d=428nm) diffraction grating in SiOx doped diamond like carbon (DLC) film. It is shown that employing a fabricated sensor chip, and using the proposed method of analysis of data, one can inspect kinetics of processes in liquids occurring in the vicinity of the grating surface. The method is based on the spectral composition analysis of polarized polychromatic light reflected from the sub-wavelength diffraction grating. The RI measurement system was tested with different model liquid analytes including 25wt.%, 50wt.% sugar water solutions, 10°C, 50°C distilled water, also Gram-positive bacteria Bacillus subtilis interaction with ion-permeable channels forming antibiotic gramicidin D and a murolytic enzyme lysozyme. Analysis of the data set of specular reflection spectra enabled us to follow the kinetics of the RI changes in the analyte with millisecond resolution. Detectable changes in the effective RI were not worse than Δn=10−4.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have