Abstract

A method was developed for the application of direct current comparator apparatus in resistance thermometry and in voltage measurements in combustion calorimetry. The performance of the measuring system was checked by combustions of benzoic acid combined with electrical calibration of the calorimeter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.