Abstract

Abstract Reliable analysis with high performance liquid chromatography (HPLC) requires purity of the eluting peak. The present work has combined the advantages of the availability of full spectral data from HPLC photodiode array UV detector and computer algorithms to perform chromatographic peak purity check. A deconvolution technique based on multicomponent analysis has been applied to the UV spectra of co-eluting components. This method employs residual error (Relative-fit-error, RFE) between predicted spectrum and analyte's spectrum to detect presence of other component or contaminant. Typical RFE values for uncontaminated chromatographic peaks of norethisterone and ethynyloestradiol range between 1 and 3, while contaminated peaks have RFE values as large as 145. A systematic increase in ‘relative-fit error’ from 1.10 to 145 was observed for peaks of norethisterone when contaminated to varied extent with ethynyloestradiol. Extent of peak overlap in chromatogram was also mapped out with this technique. ...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.