Abstract

The selection of a reversed-phase liquid chromatographic column with suitable selectivity for a particular separation is difficult if the brand name of the column is not known. A project to develop a chromatographic test procedure to characterize reversed-phase liquid chromatography C 18 columns was started earlier and resulted in a fast, simple, repeatable and reproducible test procedure using four column parameters. Here, this procedure is used to evaluate the diversity of columns originating from the same batch as well as from different batches. The determination of one of the parameters, the retention factor of 2,2′-dipyridyl, was improved and a simplified test procedure is proposed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.